Description
The Model 755VES is supplied with everything necessary for the modern laboratory including a horizontal rotor, buckets, caps, and inserts for tubes ranging from 1 – 50 mL. The exclusive airflow and rotor design guarantees the highest quality sample separations and maintains the integrity of sensitive samples.
- Large Capacity – up to twenty-four (24) 10mL test tubes, twelve (12) 15mL test tubes, six (6) 50mL tubes, 2 microplates
- Brushless-DC motor
- Full horizontal separation for best results
- Automatic braking reduces cycle time
- Precise variable speed control
- Audible end-of-run indicator
- Push-button operation/safety lid-lock
- ETL compliant to ANSI/UL standards
- Cycle counter to monitor centrifuge usage
Model 755VES: Features
- Electronic Timer with Display: Selectable from 1 to 99 minutes with hold feature
- Rotor: 24 place horizontal, 6 place horizontal for 50mL tubes, 2 position microplate
- Tube sizes: Up to twenty-four 1-10 mL tubes, up to twelve 15 mL tubes, up to six 50 mL tubes, up to two microplates
- Electronic Speed Control with Display: Set the operating speed by RPM or RCF xg
- Brushless DC Motor: Years of operation with no routine maintenance
- Programmable Memory Settings: Customize and save the run from start to finish
- Audible End-of-Run Indicator: Sounds when the rotor has come to a stop
- Quiet Operation with Low Vibration: Double-encased motor assures quiet performance and low imbalance vibration
- Automatic lid lock: Centrifuge will not run with the lid open; lid cannot be opened while samples are spinning
- Positive lid latching: Firm mechanical latching of the lid prevents the escape of debris in case of breakage or an accident
- Imbalance Detection: If an imbalance is detected, the centrifuge will come to a safe controlled stop
- Welded Steel Construction: Provides strength and durability for years of use
- Clear shatter proof lid: Allows for safe sample observation and optical calibrations
- Regulatory Approvals: ETL compliant to ANSI/UL standards
Applications
- Coagulation
- Cytology
- Platelet Poor Plasma
- Platelet Rich Plasma
- Urinalysis
- General Cell/Tissue Processing
Electrical | |
Voltage | 115 (± 10) Volts, 220/230 (± 20) Volts |
Frequency | 50/60 Hz |
Max. Power Requirements | 115 Watts for 115V |
230 Watts for 230V | |
Protection Breaker | 4 Amp. Re–settable |
Centrifuge Motor | 1/2 H.P. Brushless DC |
General | |
Overall dimensions | Height: 12.5 in. (32 cm) |
Width: 10.5 in. (27 cm) | |
Depth: 16.5 in. (42 cm) | |
Weight | 37 lbs. (17 kg) |
Acceleration time | 35 seconds with the 4 position carrier in the 6 place horizontal rotor |
31 seconds with the 6 place 50mL horizontal rotor | |
Deceleration time | 30 seconds with the 4 position carrier in the 6 place horizontal rotor |
36 seconds with the 6 place 50mL horizontal rotor | |
Timer | Digital, 1-99 minutes |
6 Place Horizontal Rotor (Standard) | Speed: 4300 (± 100) RPM |
Force: 3150 (± 90) x g with | |
6 Place 50 mL Horizontal Rotor (Optional) | Speed: 4300 (± 100) RPM |
Force: 3150 (± 90) x g |